Scilogex Humidity Chamber - Refrig. Incubator
HCN4-24
or email us at info@scilogex.com
or email us at info@scilogex.com
Product Specifications:
Performance:
HCN4-24 +/-0.5°C @ +4°C to +70°C. The published temperature variation is derived from the maximum deviation of the thermocouple located nearest the chamber geometric center during the entire test period. (i.e., +25.0°C min and +26.0°C max divided by two would be a variation of +/-0.5°C).
HCN4-24GD Temperature Variation: +/-0.5°C @ +20°C to +60°C. The published temperature variation is derived from the maximum deviation of the thermocouple located nearest the chamber geometric center during the entire test period. (i.e., +25.0°C min and +26.0°C max divided by two would be a variation of +/-0.5°C).
HCN4-52 Temperature Variation: +/-0.5°C @ +4°C to +70°C. The published temperature variation is derived from the maximum deviation of the thermocouple located nearest the chamber geometric center during the entire test period. (i.e., +25.0°C min and +26.0°C max divided by two would be a variation of +/-0.5°C).
HCN4-52GD Temperature Variation: +/-0.5°C @ +20°C to +60°C. The published temperature variation is derived from the maximum deviation of the thermocouple located nearest the chamber geometric center during the entire test period. (i.e., +25.0°C min and +26.0°C max divided by two would be a variation of +/-0.5°C).
Temperature Uniformity:
HCN4-24 +/-1°C @ +4°C to +70°C. The published temperature uniformity is derived from the maximum deviation of 9 thermocouples are placed on 3 horizontal planes, each plane having the thermocouples evenly spaced diagonally across the shelf from the left and right inner wall, and the middle sensor placed in the approximate geometric center of the shelf.
HCN4-24GD +/-1°C @ +20°C to +60°C. The published temperature uniformity is derived from the maximum deviation of 9 thermocouples are placed on 3 horizontal planes, each plane having the thermocouples evenly spaced diagonally across the shelf from the left and right inner wall, and the middle sensor placed in the approximate geometric center of the shelf.
HCN4-52 +/-1°C @ +4°C to +70°C. The published temperature uniformity is derived from the maximum deviation of 9 thermocouples are placed on 3 horizontal planes, each plane having the thermocouples evenly spaced diagonally across the shelf from the left and right inner wall, and the middle sensor placed in the approximate geometric center of the shelf.
HCN4-52GD T+/-1°C @ +20°C to +60°C. The published temperature uniformity is derived from the maximum deviation of 9 thermocouples are placed on 3 horizontal planes, each plane having the thermocouples evenly spaced diagonally across the shelf from the left and right inner wall, and the middle sensor placed in the approximate geometric center of the shelf.
Humidity Variation:
HCN4-24 +/-3% @ +4°C to +70°C (humidity-controlled range) and RH within performance graph. Humidity variation is derived from the maximum deviation of the humidity sensor during the test period.
HCN4-24GD +/-3% @ +20°C to +60°C (humidity-controlled range) and RH within performance graph. Humidity variation is derived from the maximum deviation of the humidity sensor during the test period.
HCN4-52 +/-3% @ +4°C to +70°C (humidity-controlled range) and RH within performance graph. Humidity variation is derived from the maximum deviation of the humidity sensor during the test period.
HCN4-52GD +/-3% @ +20°C to +60°C (humidity-controlled range) and RH within performance graph. Humidity variation is derived from the maximum deviation of the humidity sensor during the test period.
Additional Resources and Information:
HCN4-24
Specification Data SheetHCN4-24GD
Specification Data SheetHCN4-52
Specification Data SheetHCN4-52GD
Specification Data Sheet
Scilogex Humidity Chamber - Refrig. Incubator